Frequency Response

scpi_control.frequency_response measures a Bode plot: it steps a function generator across a list of frequencies, captures two oscilloscope channels at each step, and estimates the gain and phase of whatever sits between them. This guide covers the two-channel wiring, sweep(), what each ResponsePoint diagnostic means, autoranging, the exclusion reasons, the CSV format, and plotting -- closing with a clearly-scoped accounting of what has and has not been validated.

Wiring

A sweep needs two scope channels and one AWG channel:

  • Reference channel -- tapped straight off the generator's output, before the device under test. This is "what did we actually send."
  • Response channel -- on the device's output. This is "what came back."

sweep() drives the AWG with a sine at amplitude_vpp, sweeps its frequency, and at each step captures both channels and divides response by reference (as complex amplitudes, so both magnitude and phase come out of one division). Nothing about the DUT itself is assumed -- the transfer function is measured, not modeled.

A Minimal Sweep

from scpi_control.frequency_response import sweep

result = sweep(
    scope,
    awg,
    reference_channel=1,
    response_channel=2,
    start_hz=100.0,
    stop_hz=10_000.0,
    points_per_decade=5,
    amplitude_vpp=2.0,
)

for point in result.usable():
    print(point.frequency_hz, point.gain_db, point.phase_deg)

print(result.cutoff_hz())  # -3 dB corner, interpolated between points

start_hz/stop_hz/points_per_decade log-space the sweep for you (log_spaced_frequencies(), also exported); pass an explicit frequencies= list instead if you want particular points, e.g. densely spaced around a resonance. The AWG channel's prior state (function, frequency, amplitude, enabled) is restored when sweep() returns or raises -- the scope's state is deliberately left where the sweep ended, so the final ranging is there to inspect.

Every argument is validated before any wire traffic: a bad channel number or a non-positive frequency raises InvalidParameterError immediately. A failure mid-sweep raises FrequencySweepError, whose .partial attribute holds every point measured before the failure -- on real hardware a sweep can be a long time on the bench, and a failure at point 30 of 40 should not throw away the first 30.

See examples/frequency_response_sweep.py for a complete, runnable version of the above, including the analytic comparison used in this page's Accuracy and limits section.

Reading a ResponsePoint

Each point carries its diagnostics alongside its answer, so you don't have to take a gain or phase number on faith:

Field Meaning
frequency_hz The frequency this point asked the generator for.
gain_db Response/reference magnitude ratio, in dB. None if the point was excluded.
phase_deg Response phase relative to the reference, in degrees. None if excluded.
reference_vpp / response_vpp Raw peak-to-peak volts of each capture -- useful for spotting clipping or a flat trace at a glance.
cycles_in_window How many drive-frequency cycles the capture's time axis actually spans, measured from the returned samples (not assumed from the requested timebase). More cycles means less spectral leakage in the gain/phase estimate.
samples_per_cycle Samples per drive-frequency cycle, likewise measured. Below 20 (MIN_SAMPLES_PER_CYCLE), the point is still kept, but the library logs a warning that its phase is coarse -- worth checking samples_per_cycle yourself near the top of a sweep.
volts_per_div The response channel's vertical scale for the accepted capture -- what autoranging landed on.
excluded_reason None when gain_db is present; otherwise a string explaining why the point has no answer. The two are never both set and never both unset -- ResponsePoint enforces the pairing at construction.

A ResponsePoint is never a guess wearing an answer's clothes: if the library could not trust the numbers, gain_db/phase_deg are None and excluded_reason says why, rather than a value that happens to be wrong.

Autoranging

Real vertical resolution is finite -- 8-bit ADC codes on a legacy capture, a wider but still finite grid on a modern one -- so a response that's too small for the current scale is unreadable, and one that's too large is clipped. sweep(..., autorange=True) (the default) handles this per point:

  1. Capture both channels at the current scale.
  2. If the response's measured peak-to-peak doesn't already fill about 6 of the 8 vertical divisions, compute a better vertical scale and recapture.
  3. Only at the very first point, do the same check for the reference channel. The reference is the drive amplitude, which is constant by construction, so re-ranging it every point would spend a capture to reach the same answer every time.

Chosen scales are rounded onto the 1-2-5 sequence (0.1, 0.2, 0.5, 1.0, 2.0, ... V/div) -- the sequence real oscilloscope firmware is believed to coerce vertical (and horizontal) settings onto, though that belief itself is unverified against real firmware (see Accuracy and limits).

That makes the capture cost of a point at most two acquisitions -- both channels together each time -- and at most three at the first point: the initial capture, a response rescale if the response didn't already fill the screen, and, only at point zero, a reference rescale too. It is never three after the first point, because the reference is never re-ranged again.

Why it matters: a fixed vertical scale is wrong at both ends of a wide sweep, and the two failure modes are different in their consequences. On the RC low-pass this project measures against, holding the scale fixed cost 0.885 dB of error at 10 kHz -- a small-looking response squeezed into too few ADC codes, quietly wrong rather than absent. Earlier in this project's development, the same fixed-scale setup at 100 kHz produced a gain of -inf dB -- 20 * log10(0), from a response that had quantized flat -- returned as a bare number instead of being caught as an exclusion. That gap is exactly why ResponsePoint carries volts_per_div and why an unmeasurable point returns excluded_reason instead of a number today: running that same 100 kHz, fixed-1 V/div, autorange-disabled case now correctly returns excluded_reason="response below vertical resolution" rather than a number, because the response capture's peak-to-peak comes back at 0.0 V (quantized flat) instead of a plausible-looking level. A reader who can see the scale a point was measured at, or see that it was refused outright, has a chance to catch what a bare number never shows.

Exclusion Reasons

When a point's diagnostics say it can't be trusted, excluded_reason is one of the following. gain_db/phase_deg are None in every case.

excluded_reason What happened What to do
capture failed for channel {N} / capture failed for channels {a, b} The scope did not return usable data for that channel on that acquisition (link error, timeout, etc.). Check the connection/log for the underlying error. A single isolated occurrence may be transient; the sweep continues past it. Persistent failures on one channel usually mean it isn't actually enabled or connected.
reference below vertical resolution — source connected? The reference channel's capture is at or below the quantization floor. Check that the generator is actually connected to the reference channel and that amplitude_vpp matches what you expect the generator to output. Because the reference is only autoranged once, at the very first point, a wrong amplitude assumption here can affect the whole sweep.
response reaches beyond ±4 divisions (clipped or off screen) The response left the visible vertical range (clipped or drawn past the graticule) on the scale it was captured at, with no autorange left to try (either it's disabled, or it already tried once and the result still didn't fit). Check whether the DUT has more gain than expected, or lower amplitude_vpp. If autoranging is enabled, remember it attempts only one rescale per point, so a DUT with a very large, frequency-dependent gain swing can still outrun it at some points.
response below vertical resolution The response is too small to resolve on the scale it was captured at (again, with no further autorange available). Expected deep in a filter's stopband, far past the corner -- there may simply be very little signal left to measure. If it happens close to or before where you expect the passband, check wiring and DUT assumptions; a single autorange attempt may not be enough for a very sharp roll-off, and denser points_per_decade near the corner can help.
reference carries no energy at the drive frequency The reference passed the floor/off-screen checks, but the single-frequency estimate at the drive tone came back exactly zero. Rare; usually means the generator isn't actually outputting the requested frequency. Verify the AWG connection and that its output is enabled.
response carries no energy at the drive frequency Same as above, for the response channel. Same as above.

The CSV Format

result.to_csv("frequency_response.csv") writes a #-commented metadata header (instrument identity, firmware, library version, acquisition time, channel/settings summary, and how many of the requested points were measured), followed by plain CSV: one header row naming the same fields as ResponsePoint, then one row per point. An excluded point's numeric fields are empty, not a sentinel like -999 or nan written as text -- an empty field is unambiguous, while a sentinel can be silently misread as a real number.

That choice has a consequence worth knowing about before you reach for numpy:

  • pandas.read_csv("frequency_response.csv", comment="#") reads the file correctly, unaided. Measured, not assumed.
  • numpy.genfromtxt(..., names=True, comments="#") does NOT, even though it looks like it should. names=True always takes the first line of the input as the header, whether or not that line starts with # -- comments="#" only affects how later lines are parsed, not which line is treated as the header. Handed the raw file, genfromtxt reads the first metadata line as if it were the column names. Also measured.

Drop the # lines yourself before they reach genfromtxt -- either a filtering generator:

import numpy as np

with open("frequency_response.csv") as handle:
    rows = (line for line in handle if not line.startswith("#"))
    data = np.genfromtxt(rows, delimiter=",", names=True, dtype=None, encoding="utf-8")

or skip_header= with the exact number of metadata lines the header currently writes (fragile if that count ever changes -- the filtering generator above doesn't share that fragility).

An empty numeric field read by either library comes back as NaN, not 0 or some other plausible-looking value -- an excluded point stays visibly missing all the way through, rather than being mistaken for a real measurement of zero gain.

Plotting

result.plot(title=None) draws a Bode plot -- gain over phase, both against log frequency -- and returns the matplotlib Figure rather than showing it, so the caller decides whether to savefig(), display it, or embed it in something else:

figure = result.plot(title="RC low-pass")
figure.savefig("frequency_response.png")

Excluded points are plotted as a genuine gap, not silently dropped: the frequency axis includes every point in result.points, but an excluded point's gain and phase are plotted as NaN, which breaks the line and skips the marker at that frequency. Dropping the frequency instead would let semilogx draw one continuous line straight through the excluded region -- an interpolated claim where no measurement was made. plot() raises ValueError if there are no usable points at all (nothing to draw). See examples/frequency_response_sweep.py for a complete example that writes both the CSV and the plot.

Accuracy and limits

  • Everything measured here is against a mock instrument and an analytic model, not real hardware. The worst-case error against a closed-form first-order RC low-pass (RCLowPass, cutoff_hz=1000) over a 21-point, 100 Hz-10 kHz sweep on a 1 MSa/s mock was 0.020 dB and 1.798 degrees. No real function generator was involved in validating any of this -- there isn't one on the development bench, so nothing here has been checked against real firmware.
  • The phase worst case sits at the top of the swept range, where the mock's fixed 1 MSa/s sample rate leaves only about 100 samples per cycle at 10 kHz. A real scope raises its sample rate as the timebase shrinks, so a real instrument would very likely do better there than the mock does -- the mock's own geometry is the pessimistic case, not the measured numbers' ceiling.
  • More generally, the mock's fixed horizontal geometry shapes every accuracy figure in this section: 14 divisions, capped at 14,000 points a capture, so at 1 MSa/s the capture window never exceeds 14 ms -- only about 1.4 cycles at 100 Hz, the low end of the sweep above. That is a different mechanism from the top-end one above: a fixed point-count cap, not sample-rate-vs-timebase coupling, and nothing here establishes whether real hardware would do better or worse than the mock at the low end -- the "mock is pessimistic" argument is specific to the top of the range. Either way, these figures describe this mock's geometry, not a hardware guarantee.
  • The 1-2-5 vertical and horizontal scale coercion (round_125_up, behind both choose_timebase() and autoranging) is unverified against real firmware. The mock stores whatever scale value it's handed, so agreement between the mock and this rounding logic proves the two are internally consistent with each other, not that real instrument firmware coerces to the same values this library assumes.
  • cutoff_hz() cannot be more precise than the point spacing. It finds where the response first crosses level_db below the sweep's peak by linearly interpolating in log-frequency between the two bracketing measured points -- a real, sharper corner between those two points is invisible to it. Denser points_per_decade narrows the gap; it cannot close it to zero.
  • A capture costs at most two acquisitions per point, three at the very first point -- see Autoranging above.

See Also

  • Synthetic Signals for AwgLoopback and RCLowPass, the mock machinery this feature is built and measured on
  • Data Provenance for what the CSV metadata header's instrument/firmware/library fields actually mean and where they come from
  • examples/frequency_response_sweep.py for a runnable, hardware-free sweep against an RCLowPass DUT, including the analytic comparison